Notes for Daily Maintenance of Semi-automatic Blood Coagulation Analyzer-江苏鸿恩医疗器械有限公司
Notice
Stay away from direct sunlight and hazardous matters.
Keep the device upright. Do NOT shake or vibrate.
Keep from moisture and erosion.
Power supply voltage is 220V±10%. If power voltage is unstable, use stabilized power supply.
If fuse replacement is needed, exit the system first, unplug the power cable, and only use labeled fuse specification for replacement.(Fuse specification: 1.5A/250V/20mm)
This device is equipped with heater to maintain constant temperature of 37℃±0.5℃. Do NOT touch the heater when power is on.
To ensure test accuracy, cuvettes are disposable and not reusable.
If the battery wears out, please follow the guide in user’s manual to reenter time and date after changing battery.
Maintenance
Keep operational station clean. Do NOT drop liquid or other matters into the detector and preheating positions;
To ensure test accuracy, do NOT drop liquid and other matters in the detecting positions;
To clean the surface and keyboard, use noncorrosive detergent. Do NOT use chemical etchant;
Pay attention to the LED light in use. When orange light turns on, please use thermometer to check temperature in detector and preheating positions. If it is overheating, turn off the power immediately and investigate the cause.
Notes for Cleaning
Clean and disinfect the device after use, especially keyboard, operational station and sample injector;
Use detergent for cleaning and medical alcohol for disinfection.
Wipe the device with clean rag after disinfection.